VNA View Menus

Calibration Menu with Thru

Key Sequence: Shift, Calibrate (2) > Thru Update or Start Cal
Calibration
Start Cal: Press this soft key to open the "Next Step" List Box (see
Start Cal
Figure
step, and press Enter to measure:"
Thru Update: Press this soft key to open the selection list box (see
Thru Update
Figure 6-25 on page
"Calibration Types" on page
Back: This soft key to return to the
Back
Figure 6-23. Calibration Menu with Thru Update
Figure 6-24. Start Cal, Next Step Dialog Box
Vector Network Analyzer MG
6-24), which contains the instructions: "Connect cal component, select
6-26), then select a calibration type. Refer to Section
Next Step
Connect cal component, select step, and press Enter to measure:
Open, Port 1 (17.830 mm)
Open, Port 1 (17.830 mm)
Short, Port 2 (17.830 mm)
Short, Port 1 (17.830 mm)
Open, Port 2 (17.830 mm)
Load, Port 1
Load, Port 2
Thru, Fwd & Rev
Isolation, Fwd & Rev (optional)
Calculate and Finish Cal
Exit to Setup
Abort Cal
PN: 10580-00289 Rev. K
6-37.
"Calibration Menu" on page
6-4 Calibration Menus
6-19.
6-25