6-4 Calibration Menus
Cal Type Menu
This menu offers a choice between Standard, which is the default type, and Flex. When you
choose Flex, the instrument performs a calibration over the entire frequency range of the
instrument. You can then set any frequency range or any number of points (or both), and the
instrument interpolates the calibration coefficients and applies the calibration to the new
range or points (or both).
Example of Calibration Steps for S
Open, Port 1 (17.830 mm)
Short, Port 1 (17.830 mm)
Load, Port 1
Calculate and Finish Cal
Exit to Setup
Abort Cal
Examples of Calibration Types:
Full 2 Port – (S
Full S
– Port 1 (S
11
Response S
11
1P2P S
, S
11
21
1P2P S
, S
22
12
6-20
:
11
, S
, S
, S
)
11
21
12
22
)
11
& S
– Refl Resp – Both Ports (S
22
– 1 Path 2 Port Forward Path (S
– 1 Path 2 Port Reverse Path (S
PN: 10580-00289 Rev. K
, S
)
11
22
, S
)
11
21
, S
)
22
12
Vector Network Analyzer MG
VNA View Menus