Run/Hold
Run Hold: Press this soft key to toggle the sweep to Run or Hold.
Sweep
Sweep Type
Run/Hold
Single Cont: Press this soft key to toggle the sweep type to Single or
Run
Hold
Continuous. Single sweep sets the VNA Master to make a single sweep and
then wait for additional commands. The Continuous mode sweeps
Sweep Type
continuously.
Single
Cont
Data Points: Press this soft key to set the number of data points from
2 to 4001.
Data Points
IFBW: Press this soft key to open a list box and set the Intermediate
##
Frequency bandwidth. The default is 10 kHz. Select 10 Hz for the maximum
dynamic range, and select 100 kHz for the maximum speed.
IFBW
Sweep Averaging: Press this soft key to set the number of sweeps to use for
## Hz
averaging. The minimum number is 1.
Sweep Averaging
Configure Ports: Press this soft key to open the
#
Dithering
On Off: Press this soft key to toggle On and Off the Dithering function.
When a signal is dithered, a mathematical process removes the harmonics or
Configure Ports
other undesirable distortions, and replaces these distortions with a constant,
fixed noise level. The result is a better representation of the signal.
Dithering
RF Pwr in Hold
On Off: Press this soft key to toggle On and Off the condition of the RF
On
Off
power transmitted from Port 1 and Port 2 when the instrument is put in Hold
RF Pwr in Hold
mode. The RF power can be set either to stay on or to be turned off during
Hold. The default setting is for the RF to stay on during Hold, which helps to
On
Off
stabilize the instrument temperature.
Figure 6-52. Sweep Menu
To access the Sweep Menu, press the Sweep function hard key or press the Shift key then the