VNA View Menus

Setup Domain Menu

Key Sequence: Freq/Time/Dist > Domain Setup
Setup Frequency: Press this soft key to open the Frequency menu and to
use those soft keys to set start, stop, and center frequencies and frequency
Setup Domain
span. Choosing Setup Frequency sets the domain (x-axis) of the active trace
to Frequency (just as the
Setup
Frequency
Menu" on page
Setup Time: Press this soft key to open the
Setup
and stop times, the start and stop frequencies, windowing, and gating.
Time
Choosing Setup Time sets the domain (x-axis) of the active trace to Time (just
as the
Setup
Setup Distance: Press this soft key to open the
Distance
to set the Start and Stop distances, the start and stop frequencies, windowing,
and gating. The Start and Stop distances are set in meters (m),
centimeters (cm), or millimeters (mm). If the Application Options menu is
Setup
FGT
used to set units in feet rather than meters, the distance settings are in feet (ft)
only. Choosing Setup Distance sets the domain (x-axis) of the active trace to
Distance (just as the
Setup FGT: Press this soft key to open the Frequency Gated by Time (FGT)
menu and to use those soft keys to set start, stop, and center frequencies,
and to set frequency span and gating. Choosing Setup FGT sets the domain
Back
(x-axis) of the active trace to Frequency (just as the
key does in the Measure menu). Note that the frequency data are not the
measured frequency data (as in the Setup Frequency case), but they are time
or distance domain data that are gated and converted back into the frequency
domain. When the trace is in the Time domain and then set to FGT, the
frequency data are generated by converting the gated time domain data.
When the trace is in the Distance domain and then set to FGT, the frequency
data are generated by converting the gated distance domain data. If the trace
was set to Frequency and then converted to FGT directly, then the instrument
defaults to converting the gated distance domain data.
Back: Press this soft key to return to the previous menu.
Figure 6-3.
Setup Domain Menu
Vector Network Analyzer MG
"Domain Selection"
6-46).
"Domain Selection"
soft key does in the Measure menu).
"Domain Selection"
PN: 10580-00289 Rev. K
6-3 Domain Setup Menus
soft key does in the
"Time Menu"
and to set the start
"Distance Setup Menu"
soft key does in the Measure menu).
"Domain Selection"
"Measure
and
soft
6-5