Vector Voltmeter, Option 15
7. As shown in
Figure
in a new reference window above the table and updates the REL Amp and REL Phase
columns to display the difference between the current measurement and the saved
reference measurement.
Figure 11-7. Vector Voltmeter Relative Measurement Table – MS20xxC VNA Master
8. Before disconnecting the first DUT, save the results in the current row by using the
Select Cable soft key to move to another row in the table. When selecting a new cable,
the Vector Network Analyzer saves the results and updates measurements for the next
cable in the table.
Be sure to keep the DUT connected until a new cable has been selected in
Note
order to ensure that the proper values are saved in the table for the final report.
9. Connect the second DUT and view the difference between the first and second DUT
measurements. Be careful to avoid selecting another cable when you are done with the
measurement. Saved data can be overwritten.
10. To create a new reference, press the Clear Reference soft key followed by the Save New
Reference soft key. Consider the possibility of changing the reference by using
Select Cable to choose cable 1 again.
This completes the procedure for relative measurements using the table display.
Vector Network Analyzer MG
11-8 Measurements Using Comparison Table Display
11-7, the Vector Network Analyzer saves the current measurement
PN: 10580-00289 Rev. K
11-13