Calibration Menu - Anritsu MS2024B Manual

Vector network analyzer for anritsu rf and microwave handheld instruments
Hide thumbs Also See for MS2024B:
Table of Contents

Advertisement

VNA View Menus

Calibration Menu

Key Sequence: Shift, Calibrate (2)
Start Cal: Press this soft key to open the "Next Step" List Box, which contains
the instructions: "Connect cal component, select step, and press Enter to
Calibration
measure:" See
instruments, this key changes to
Start Cal
calibration includes a Thru measurement. See the soft key image at the
bottom of this list.
Cal Type
Cal Type: Press this soft key to open the selection list box, then select a
calibration type. A soft key offers a toggled selection of Flex or Standard.
Full 2 Port
Refer to
Cal Method
on page
SOLT
Cal Line Setup
Cal Method: Press this soft key to open the selection list box, then select a
calibration method. See
DUT Port Setup
is displayed only for MS20xxC instruments..
Cal Line Setup: Press this soft key to open the
Menu" on page
Interpolation
Menu" on page 6-27
On
Off
soft key is displayed only for MS20xxC instruments.
DUT Port Setup: Press this soft key to open the
Existing Cal Info
(Coax)" on page 6-29
on page 6-33
Cal Correction
Interpolation
On Off: Press this soft key to toggle interpolation On or Off. In Cal Type
On
Off
Flex mode, Interpolation is always On. After performing a Standard
calibration, you can turn Interpolation On and then change the frequency
range (smaller) or reduce the number of data points. Refer to
on page
Existing Cal Info: Press this soft key to open the "Existing Calibration
Information" List Box. Press Enter or Esc to close the list box.
Cal Correction
On Off: Press this soft key to toggle calibration correction On and Off.
Thru Update or Start Cal: This soft key is displayed in place of the Start Cal
soft key when a completed calibration includes a Thru measurement.
Thru Update
or
Pressing this key opens the
Start Cal
offers a Thru update or a new calibration. This soft key is displayed only for
MS20xxC instruments.
Figure 6-18. Calibration Menu
Vector Network Analyzer MG
Figure 6-24 on page 6-25
"Cal Type
Menu". Also refer to Section
6-37. See
Figure 6-19 on page 6-21
Figure 6-22 on page 6-24
6-27. A soft key (Cal Line Type) in the
offers a toggled selection of Coax or Waveguide. This
or the
"DUT Port Setup Menu (Waveguide)"
if the Cal Line Type is set to Waveguide..
6-22.
"Calibration Menu with Thru" on page
PN: 10580-00289 Rev. K
for examples. Only for MS20xxC
"Thru Update or Start Cal"
"Calibration Types"
for Cal Type examples.
Cal Type
Flex
Standard
for examples. This soft key
"Calibration Line Setup
"DUT Port Setup Menu
6-4 Calibration Menus
when a completed
"Calibration Line Setup
"Interpolation"
6-25, which
6-19