High Band Verification
(> 50 GHz)
NOTE
NOTE
NOTE
Performance Verification
Verification Procedures
Before verification of the high band frequencies it is necessary to: 1) select
new hardware for the system and 2) perform a calibration (for the frequency
band > 50 GHz).
1. From the Main Menu select
2. Select the proper hardware setup for High Band Verification (> 50 GHz)
using
Table 4-2 on page 4-4
Table 4-3 on page 4-4
3. Select
and
Done
Prior Menu
4. Select
to go to the System Performance Verification Menu.
Verify System
5. Select
and follow the program prompts to calibrate the system
System Cal
for frequencies >50 GHz. See
required and suggested calibration storage locations (refer to
page
4-9).
6. Select
to go to the Standard Selection Menu.
Select Standard
7. To verify the system use the Verification Standards provided with the
85059A 1.0 mm Precision Calibration and Verification Kit (refer to
Table 4-6 on page
4-10). Using the menu select the verification standard
and enter the standard's serial number. The verification standards may
be measured in any order.
It is recommended to connect the Verification Standards to the cable on
port 2 prior to attempting to connect to port 1. Then follow the
recommended procedure for making a thru connection outlined under
on page 4-9
to avoid damaging the cable.
8. Change default for
Cal Set for Broadband Measurement
It is suggested that the High Band Calibration be stored in Cal Set 3 (Refer
to
Table 4-1 on page
4-3).
9. Verify
Cal Set for 45 MHz Measurement (If Required)
10. Select
and follow the program prompts.
Done
11. Select
to measure the verification standards. Follow the
Measure Data
program prompts.
4-12 8510XF Network Analyzer Systems
then
System Config
Hardware Config
for the E7352A 110 GHz System or
for the E7342A 85 GHz System.
to return to Main Menu.
Table 4-1 on page 4-3
is set to
.
for the calibrations
step 3 on
step 3
from
to
.
1
3
.
No

Advertisement

Table of Contents
loading

This manual is also suitable for:

E7340aE7342aE7350aE7352a

Table of Contents