Replacement Procedure

A11 Source Module Replacement

A11 Source Module Replacement

Tools Required

Removal Procedure

Refer to Figure 4-13 for this procedure.
Step 1. Remove the Outer Cover as described in "Outer Cover Replacement" on page 54.
Step 2. Remove the Front Panel as described in "Front Panel Replacement" on page 56.
Step 3. Remove the A3/A4 Receiver Module as described in "A3/A4 Receiver Module
Replacement" on page 76.
Step 4. Remove the two TORX T10 screws (item 1) fastening the A11 Source Module.
Step 5. Slide forward the A11 Source Module.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws (item 2, and
item 3).
80
torque screwdriver, T15 and T20
torque screwdriver, T10 (set to 9 in-lb)
open torque wrench, 5/16 inch
flat edge screwdriver
Chapter 4