Blanking Test - Datalogic Magellan 9800i Product Reference Manual

Hide thumbs Also See for Magellan 9800i:
Table of Contents

Advertisement

Increasing- Load Test (Single Interval Phase 2)
(Dual Interval Phase 3)
After completing the Shift Test, you must complete the Increasing Load Test using 20.0, 25.0
and 30.0 pounds of weight.
SINGLE INTERVAL
After completing the Shift Test, you must complete the
Increasing Load Test using 20.0, 25.0 and 30.0 pounds
of weight.
1. With the scale starting at zero, place 20.0 pounds
in the center of the weighing surface and check
that the display reads between 19.99 and 20.01
lb.
2. Place an additional five pounds in the center of
the weighing surface, increasing the load to 25.0
pounds and check that the display reads between
24.99 and 25.01 lb.
3. Place an additional five pounds in the center of
the weighing surface increasing the load to 30.0
pounds and check that the display reads between
29.99 and 30.01 lb.
4. Remove the weights and verify that the display
reads 0.00.
5. You have completed the Increasing-Load Test
(Phase 2)

Blanking Test

This test ensures that the scanner/scale will indicate its weighing capability has been exceeded if
a weight greater than 0.05 pounds over its maximum upper weight is placed upon the unit.
1. Place weights that total the upper weight limit plus 0.051 pounds on the scale. For exam-
2. Verify that the display shows a dash and three hyphens (_ - - -). This is the overweight
Product Reference Guide
The upper limit of the scale is configurable according to the POS interface
type and may not necessarily be set at 30 pounds, which is the standard
setting. For this test, continue to place weights in five pound increments
only up to the upper weight limit set for your scale.
NOTE
ple: If the upper limit is set at 30 pounds, place weights totaling 30.051 pounds on the
scale.
indication.
The scanner/scale may blank (show an underscore and three hyphens) at any
weight greater than its upper weight limit, but must blank when the upper
weight limit plus 0.051 pounds are set upon the weighing surface.
NOTE
DUAL INTERVAL
After completing the Shift Test, you must complete the
Increasing Load Test using 8.0, 10.0 and 12.0 pounds
of weight.
1. With the scale starting at zero, place 20.0 pounds
in the center of the weighing surface and check
that the display reads between 19.99 and 20.01
lb.
2. Place an additional five pounds (25 pounds total)
in the center of the weighing surface. Check that
the display reads between 24.99 and 25.01 lb.
3.
Place an additional five pounds in the center of
the weighing surface increasing the load to 30.0
pounds and check that the display reads between
29.99 and 30.01 lb.
4. Remove the weights and verify that the display
reads 0.00.
5. You have completed the Increasing-Load Test
(Phase 3).