Restore Default Setup; Viewing Status - Teledyne HDO9000 Operator's Manual

High definition oscilloscopes
Table of Contents

Advertisement

Note: You will undo all new measurements or math function definitions entered since the Auto
Setup when you Undo the Auto Setup. Perform this work when the instrument is not in the
Auto Setup mode if you wish for it to persist.

Restore Default Setup

Restore the factory default state by pressing the front panel Default Setup button. You can also
restore default settings by choosing File > Recall Setup > Recall Default.
Setting
Channel/Vertical
Timebase
Trigger
Display
Cursors
Measurements
Math

Viewing Status

All instrument settings can be viewed through the various Status dialogs. These show all existing
acquisition, trigger, channel, math function, measurement and parameter configurations, as well as
which are currently active.
Access the Status dialogs by choosing the Status option from the Vertical, Timebase, Math, or
Analysis menus (e.g., Channel Status, Acquisition Status).
Default
C1 on at 50 mV/div Scale, 0 V Offset
HD Mode on, Real Time Sampling at 50 ns/div, 0 Delay, 20 kS at 40
GS/s, 2.5 MS Memory
C1 with an Auto Positive Edge, DC Coupling, 0 V Level
Auto Grid
Off
Cleared
Cleared
Acquisition
37

Advertisement

Table of Contents
loading

Table of Contents