Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault.
Status signal
Example
12.4.2
Calling up remedy information
Remedy information is provided for every diagnostic event to ensure that problems can be
rectified quickly:
• On the home page
Remedy information is displayed in a separate field below the diagnostics information.
• In the Diagnostics menu
Remedy information can be called up in the working area of the user interface.
The user is in the Diagnostics menu.
1. Call up the desired parameter.
2. On the right in the working area, mouse over the parameter.
A tool tip with remedy information for the diagnostic event appears.
12.5
Adapting the diagnostic information
12.5.1
Adapting the diagnostic behavior
Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagnostic behavior submenu.
Expert → System → Diagnostic handling → Diagnostic behavior
Available diagnostic behaviors
The following diagnostic behaviors can be assigned:
Diagnostic behavior
Alarm
Warning
Logbook entry only
Off
Diagnostic information
Diagnostic code
Diagnostic number
↓
↓
842
A0013958
3-digit number
Description
The device stops measurement. The totalizers assume the defined alarm
condition. A diagnostic message is generated.
The device continues to measure. Measured value output via PROFINET and
totalizers are not affected. A diagnostic message is generated.
The device continues to measure. The diagnostic message is only displayed in
the Event logbook submenu (Event list submenu) and is not displayed in
alternating sequence with the operational display.
The diagnostic event is ignored, and no diagnostic message is generated or
entered.
Diagnostics and troubleshooting
Short text
↓
Process limit
93