Diagnostics and troubleshooting
Diagnostic information
No.
410 Data transfer
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
412 Processing download
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
437 Configuration incompatible
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
106
12.6.3
Diagnostic of configuration
Short text
Bad
Maintenance alarm
0x24 to 0x27
F
Alarm
Short text
Uncertain
Initial value
0x4C to 0x4F
C
Warning
Short text
Bad
Maintenance alarm
0x24 to 0x27
F
Alarm
Remedy instructions
1. Check connection
2. Retry data transfer
Remedy instructions
Download active, please wait
Remedy instructions
1. Restart device
2. Contact service
Proline Promass A 100 PROFINET
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Status
• Volume flow