Commissioning
Parameter
Corrected volume flow calculation
External reference density
Fixed reference density
72
Navigation
"Setup" menu → Advanced setup → Calculated values
‣
Calculated values
‣
Corrected volume flow calculation
"Corrected volume flow calculation" submenu
Navigation
"Setup" menu → Advanced setup → Calculated values → Corrected volume flow calculation
‣
Corrected volume flow calculation
Corrected volume flow calculation
(1812)
External reference density (6198)
Fixed reference density (1814)
Reference temperature (1816)
Linear expansion coefficient (1817)
Square expansion coefficient (1818)
Parameter overview with brief description
Prerequisite
–
In the Corrected volume flow
calculation parameter, the
External reference density
option is selected.
The Fixed reference density
option is selected in the
Corrected volume flow
calculation parameter
parameter.
Description
Select reference density for
calculating the corrected
volume flow.
Shows external reference
density.
Enter fixed value for reference
density.
Proline Promass A 100 PROFINET
→ 72
→ 72
→ 72
→ 72
→ 73
→ 73
→ 73
Selection / User
Factory setting
interface / User
entry
• Fixed reference
–
density
• Calculated
reference density
• Reference density
by API table 53
• External reference
density
Floating point
–
number with sign
Positive floating-
–
point number
Endress+Hauser