Proline Promass A 100 PROFINET
Diagnostic information
No.
438 Dataset
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
453 Flow override
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
484 Simulation Failure Mode
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Endress+Hauser
Short text
Uncertain
Maintenance demanded
0x68 to 0x6B
M
Warning
Short text
Good
Function check
0xBC to 0xBF
C
Warning
Short text
Bad
Function check
0x3C to 0x3F
C
Alarm
Remedy instructions
1. Check data set file
2. Check device configuration
3. Up- and download new configuration
Remedy instructions
Deactivate flow override
Remedy instructions
Deactivate simulation
Diagnostics and troubleshooting
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Status
• Volume flow
107
Need help?
Do you have a question about the Proline Promass A 100 PROFINET and is the answer not in the manual?