Proline Promass A 100 PROFINET
Endress+Hauser
2
1
Status area with status signal
2
Diagnostic information → 91
3
Remedial measures with service ID
In addition, diagnostic events which have occurred can be shown in the Diagnostics
menu:
• Via parameter → 117
• Via submenu → 118
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Symbol
Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).
Out of specification
The device is being operated:
Outside its technical specification limits (e.g. outside the process temperature range)
Maintenance required
Maintenance is required. The measured value remains valid.
The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107.
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault.
Diagnostics and troubleshooting
1
3
A0031056
91