Proline Promass A 100 PROFINET
Diagnostic information
No.
830 Sensor temperature too high
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
831 Sensor temperature too low
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
832 Electronic temperature too high
Measured variable status [from the factory]
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Endress+Hauser
Short text
Uncertain
Process related
0x78 to 0x7B
S
Warning
Short text
Uncertain
Process related
0x78 to 0x7B
S
Warning
Short text
1)
Good
Ok
0x80 to 0x83
S
Warning
Remedy instructions
Reduce ambient temp. around the sensor
housing
Remedy instructions
Increase ambient temp. around the sensor
housing
Remedy instructions
Reduce ambient temperature
Diagnostics and troubleshooting
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temperature
• Volume flow