Diagnostics and troubleshooting
Diagnostic information
No.
311 Electronic failure
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
382 Data storage
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
383 Memory content
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
104
Short text
Bad
Maintenance alarm
0x24 to 0x27
M
Warning
Short text
Bad
Maintenance alarm
0x24 to 0x27
F
Alarm
Short text
Bad
Maintenance alarm
0x24 to 0x27
F
Alarm
Remedy instructions
1. Do not reset device
2. Contact service
Remedy instructions
1. Insert DAT module
2. Change DAT module
Remedy instructions
1. Restart device
2. Check or change DAT module 3. Contact
service
Proline Promass A 100 PROFINET
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Status
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Status
• Volume flow
Endress+Hauser
Need help?
Do you have a question about the Proline Promass A 100 PROFINET and is the answer not in the manual?