Endress+Hauser Proline Promass A 100 PROFINET Operating Instructions Manual page 119

Coriolis flowmeter
Hide thumbs Also See for Proline Promass A 100 PROFINET:

Advertisement

Proline Promass A 100 PROFINET
Endress+Hauser
In addition to the operating time when the event occurred, each event is also assigned a
symbol that indicates whether the event has occurred or is finished:
• Diagnostics event
• : Occurrence of the event
• : End of the event
• Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
• Via web browser →  92
• Via "FieldCare" operating tool →  93
• Via "DeviceCare" operating tool →  93
For filtering the displayed event messages →  119
12.9.2
Filtering the event logbook
Using the Filter options parameter you can define which category of event message is
displayed in the Events list submenu.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)
12.9.3
Overview of information events
Unlike a diagnostic event, an information event is displayed in the event logbook only and
not in the diagnostic list.
Info number
I1000
--------(Device ok)
I1089
Power on
I1090
Configuration reset
I1091
Configuration changed
I1111
Density adjust failure
I1137
Electronic changed
I1151
History reset
I1155
Reset electronic temperature
I1157
Memory error event list
I1185
Display backup done
I1186
Restore via display done
I1187
Settings downloaded with display
I1188
Display data cleared
I1189
Backup compared
I1209
Density adjustment ok
I1221
Zero point adjust failure
I1222
Zero point adjustment ok
Diagnostics and troubleshooting
Info name
119