Proline Promass A 100 PROFINET
Diagnostic information
No.
842 Process limit
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
843 Process limit
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Diagnostic information
No.
862 Partly filled pipe
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Endress+Hauser
Short text
Good
Ok
0x80 to 0x83
S
Warning
Short text
Good
Ok
0x80 to 0x83
S
Warning
Short text
Uncertain
Process related
0x78 to 0x7B
S
Warning
Remedy instructions
Low flow cut off active!
1. Check low flow cut off configuration
Remedy instructions
Check process conditions
Remedy instructions
1. Check for gas in process
2. Adjust detection limits
Diagnostics and troubleshooting
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
Influenced measured
variables
• Carrier mass flow
• Concentration
• Density
• Dynamic viscosity
• Kinematic viscosity
• Mass flow
• Sensor integrity
• Reference density
• Corrected volume flow
• Target mass flow
• Temp. compensated
dynamic viscosity
• Temp. compensated
kinematic viscosity
• Temperature
• Volume flow
113