Note:
To disassemble the filament housing assembly and replace a filament, see
Replacing the inner source
7.
Lift the ionization chamber and the repeller assembly away from the inner source.
Figure 4–16: Removing the ionization chamber and repeller from the inner source
8.
Remove the repeller and the repeller rod from the repeller holder.
Figure 4–17: Removing the repeller and repeller rod from the repeller holder
Repeller
Repeller holder
Repeller rod
9.
Clean the repeller and the ionization chamber (see
chamber).
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
filament.