Changing The Inner Source Assembly From Ci To Ei Operation - Waters Xevo TQ-GC Overview And Maintenance Manual

Table of Contents

Advertisement

Note:
Ensure that the CI filament housing assembly is fitted the correct way.
Figure 3–8: Correctly fitted CI filament housing assembly
CI filament housing assembly fitted with one uncoated rhenium CI filament
13. Fit the inner source assembly to the instrument (see
the
instrument).
14. Start the instrument (see
Important:
ensure that you select Current Control from the Tune window.
3.2 Changing the inner source assembly from CI to EI
operation
Required materials:
• Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Warning:
Warning:
enclosure open.
Preparing to start the mass
When you manually tune the instrument for CI+ or CI- operation in MassLynx,
To avoid eye injury, use eye protection when performing this procedure.
To avoid burn injuries, take great care while working with the source
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Fitting the inner source assembly to
spectrometer).
Page 46