4.6.2 Removing the mounting flange assembly from the instrument

Required materials:
• Chemical-resistant, powder-free gloves
• Flat-blade screwdriver
To remove the mounting flange assembly:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Warning:
Warning:
enclosure open.
Warning:
this component can be hot. Wait until the mounting flange has sufficiently cooled before
you handle it.
1.
Prepare the instrument so you can work safely on the source (see
instrument for working on the
2.
Ensure that all gases are turned off.
Cleaning the inner source components
To avoid eye injury, use eye protection when performing this procedure.
To avoid burn injuries, take great care while working with the source
To avoid burn injuries, take great care when you handle the mounting flange;
source).
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
and
Replacing the inner source
Page 58
filament.