Figure 4–13: Standing the inner source upright
4.
Allow the source to cool.
5.
Holding the filament housing assembly, push it down and then rotate it approximately 30°
in a counter-clockwise direction.
Figure 4–14: Rotating the filament housing assembly
6.
Lift the filament housing assembly away from the inner source assembly.
Figure 4–15: Removing the filament housing assembly from the inner source
assembly