Figure 4–6: Outer source assembly source block screws
Source block screw x 3
7.
With the screws loosened, remove the outer source assembly from the instrument.
Note:
The outer source is still attached to the PCB feed-through cable.
8.
Disconnect the 10-pin PCB feed-through connector from the PCB holder on the outer
source.
Figure 4–7: Disconnecting the 10-pin PCB feed-through connector
10-pin PCB feed-through connector
16-pin source-housing vacuum feed-through connector
PCB feed-through cable (supplies power to the PCB)
See also
Cleaning the outer source lens
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
stack.