12. Close the column-compartment drawer, and push downward on the
handle, to secure the latch.
Drawer handle
13. Close the outer door of the CM-30S.
Replacing the ferrule on the column fitting
Replace the ferrule on the column fitting if the column is leaking.
Warning:
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column compartment's
internal temperature to ensure all components are cool.
Warning:
when replacing the ferrule on the column fitting. Know the physical and
chemical properties of the solvents used (see the Material Safety Data
Sheets for the solvents in use).
Notice:
chemical-resistant, powder-free gloves when replacing the ferrule on the
column fitting.
To prevent burn injuries, set the column temperature to Off,
To prevent injury, always observe Good Laboratory Practice
To prevent contaminating system components, wear clean,
Replacing the ferrule on the column fitting
October 7, 2014, 715004178 Rev. B
119