Waters Xevo TQ-GC Overview And Maintenance Manual page 60

Table of Contents

Advertisement

Warning:
To avoid burn injuries, take great care when you handle the outer source;
this component can be hot. Wait until the outer source has sufficiently cooled before you
handle it.
1.
Prepare the instrument so you can work safely on the source (see
instrument for working on the
2.
Ensure that all gases are turned off.
3.
Remove the inner source assembly from the instrument (see
assembly).
4.
Remove the mounting flange assembly from the instrument (see
flange
assembly).
Result:
The outer source assembly is now accessible.
Figure 4–5: Outer source assembly fitted to the mass spectrometer
Outer source assembly
PCB feed-through cable (supplies power to the PCB)
16-pin source-housing vacuum feed-through connector
Outer source captive screw x 2
5.
To avoid shearing the column, loosen the ferrule interface in the GC oven and retract the
column approximately 4 cm through the heated transfer line to allow the outer source to be
removed. For instructions on how to do this, refer to the documentation included with the
GC oven, observing all cautions and warnings.
6.
With the column retracted, face the instrument from the front, hold the outer source with
one hand, and then, using the wrench, loosen the two outer source captive screws.
Attention:
secure the source block to the outer source assembly.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
source).
Do not, at this time, loosen the three source block screws that
Page 60
Preparing the
Removing the inner source
Removing the mounting