Fitting The Inner Source Assembly To The Instrument - Waters Xevo TQ-GC Overview And Maintenance Manual

Table of Contents

Advertisement

1.
Ensure that the outer source assembly is fitted (see
the
instrument).
2.
Align the mounting flange with the aperture on the instrument.
3.
Hold the mounting flange against the instrument, and then tighten the four thumb screws.
Figure 4–10: Tightening the mounting flange assembly's thumb screws
Result:
4.
To fit the inner source to the instrument, see

4.6.6 Fitting the inner source assembly to the instrument

Required materials:
• Chemical-resistant, powder-free gloves
To fit the inner source assembly:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
1.
Ensure that the outer source assembly is fitted (see
the
instrument).
2.
Ensure that the mounting flange is fitted (see
instrument).
3.
Align the guidance prongs on the inner source assembly with the grooves in the mounting
flange.
The inner source assembly is ready to be fitted to the instrument.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Fitting the outer source assembly to
Fitting the inner source to the
Fitting the outer source assembly to
Fitting the mounting flange assembly to the
Page 64
instrument.