Figure 1–19: Product ion spectrum
MS1–Static (at precursor mass)
Collision cell–Fragment precursor ions and pass all masses
MS2–Scanning
1.9.1.1 Typical applications
You typically use product ion spectrum for the following applications:
• Method development for MRM screening studies:
• Identifying product ions for use in MRM transitions.
• Optimizing CID tuning conditions to maximize the yield of a specific product ion to be used
in MRM analysis.
• Structural elucidation.

1.9.2 Precursor (parent) ion spectrum

Figure 1–20: Precursor ion spectrum
MS1–Scanning
Collision cell–Fragment precursor ions and pass all masses
MS2–Static (at product mass)
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Page 32