Warning:
To avoid burn injuries, take great care when you handle the inner source;
this component can be hot. Wait until the inner source has sufficiently cooled before you
handle it.
Warning:
To prevent the inner source from toppling over and causing injury, stand it
upright on its handle on a flat bench.
Important:
Ensure that you use the EI filament housing assembly fitted with two yttrium oxide-
coated rhenium EI filaments, and the EI ionization chamber.
To change from CI to EI operation:
1.
Prepare the instrument so you can work safely on the source (see
instrument for working on the
2.
Remove the inner source assembly from the instrument (see
assembly from the
3.
Stand the inner source upright on its handle on a flat surface.
Figure 3–9: Inner source standing upright
4.
Allow the source to cool.
5.
To steady the inner source, hold the handle with one hand.
6.
With the other hand, hold the CI filament housing assembly, push it down and then rotate it
approximately 30° in a counter-clockwise direction.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
source).
instrument).
Page 47
Preparing the
Removing the inner source