To replace the filament:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
Warning:
To avoid eye injury, use eye protection when performing this procedure.
Warning:
To prevent the inner source from toppling over and causing injury, stand it
upright on its handle on a flat bench.
1.
Prepare the instrument so you can work safely on the source (see
instrument for working on the
2.
Remove the inner source assembly from the instrument (see
assembly).
3.
Stand the inner source upright on its handle on a flat surface.
Figure 4–24: Inner source standing upright
4.
Allow the source to cool.
5.
Holding the filament housing assembly, push it down and then rotate it approximately 30°
in a counter-clockwise direction.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
source).
Page 74
Preparing the
Removing the inner source