Attention:
instrument's performance will deteriorate if the magnets are refitted incorrectly
into the magnet holders.
5.
Using the wrench, loosen and remove the two screws that secure the lenses and the
spacers.
Figure 4–38: Removing the screws securing the lenses and spacers
Lens securing screw x 2
6.
Remove the lenses and the PEEK spacers, and record the numbers that are etched on
them to aid reassembly.
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Do not remove the magnets from the magnet holders. The
Page 83