Figure 3–5: Aligning the CI ionization chamber
10. Lower the ionization chamber onto the repeller assembly.
Note:
The ionization chamber is asymmetrical and can only be fitted one way.
11. Align the guidance pegs on the CI filament housing assembly with the bayonet mount on
the inner source assembly.
Figure 3–6: Aligning the CI filament housing assembly
12. Lower the CI filament housing assembly into the bayonet mount, push it down and then
rotate it approximately 30° in a clockwise direction.
Figure 3–7: Securing the CI filament housing assembly