Servicing
Nephelometer and Beta Attenuation Servicing

Firmware Update

7-4
Model 5030 SHARP Monitor Instruction Manual
Firmware for the Model 5030 SHARP can be updated by exchanging its
FLASH-EEPROM. However, it is more easily accomplished through the
transmission of a new program code via the COM1 serial interface. During
transmission the program is saved in RAM. After the transmission has been
completed without corruption, the program code is written into the Flash -
EEPROM.
A download of the Parameter List (Print Format 8) should be done prior to
any firmware updates. Refer to "Print Formats" in the "Operation" chapter.
Use the following requirements and steps to update to a new firmware
version.
Requirements:
PC with Windows 98 or higher with COM1 serial interface (9 pin)
Model 5030 SHARP with jumpered X30/Pin 1-2 (insert: longest
distance from the Flash - EEPROM)
Communications Cable with the configuration shown in
This cable connects the PC 9-pin COM1 port to the Model 5030
SHARP 25-pin COM1 port.
Comm unications Cable W iring Diagram
2
3
5
6
7
4
8
Female 9-pin
Cable Connector
Connects to:
M ale 9-Pin
PC COM 1 Connector

Figure 7-3. Communications Cable Wiring Diagram and Serial Connection

2
3
7
M ale 25-Pin
Cable Connector
Connects to:
Fem ale 25-Pin
M odel 5030 COM 1 Connector
Figure
7-3.