Bench Acceptance Test - Thermo Scientific 5030 Instruction Manual

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Bench Acceptance
Thermo Fisher Scientific
Prior to installing the Model 5030, you should perform the acceptance
Test
testing procedures. These tests are conducted to evaluate the out-of-box
performance of the instrument, perform any necessary calibrations prior to
final site installation, and familiarize the user with the menu structure.
Use the following procedure to perform the bench acceptance test.
Equipment Required:
Model 5030 SHARP Monitor,
Model 5030 Power cord
Mini-ambient/heater temperature cable assembly (DO NOT USE THE
HEATED SAMPLE TUBE ASSEMBLY AT THIS POINT)
Vacuum Pump Assembly
Flow Adapter Assembly
NIST-traceable Thermometer
NIST-traceable Hygrometer
NIST-traceable Barometer
NIST-traceable Volumetric Flow Transfer Standard
1. After unpacking the instrument place the Model 5030 and accessories
onto a table or bench located within a stable indoor environment.
Prior to continuing with acceptance testing, the instrumentation and
accessories should have sufficient time to equilibrate to room
temperature due to temperature variations during shipping and/or
storage.
2. Following a sufficient equilibration period, connect the power cord to
the rear of the instrument and to a properly grounded power supply. If
the Model 5030 turns ON at this point, set the power switch to the
"OFF" position and then proceed.
3. Connect the braided vacuum hose to the top brass port on the pump
and the other end to the vacuum pump hose connection on the rear of
the monitor.
Model 5030 SHARP Monitor Instruction Manual
Installation

Bench Acceptance Test

2-7