Specifications
Thermo Fisher Scientific
Table 1–1
lists the instrument's features and specifications.
Table 1–1. Model 5030 Specifications
Feature
Concentration ranges
Minimum detectable
concentration limit
Hourly precision
Measurement time resolution
Precision between two
monitors
Span drift
Display resolution
Accuracy
Sources
Detectors
Air flow rate
Output
Analog output
Operating environment
Power supply instrument
Pump
Dimensions Instrument
Pump
Description
3
0 to 1,000 μg/m
and 0 to 10,000 μg/m
3
< 0.5 μg/m
@ 2σ (one-hour time resolution)
3
< 0.2 μg/m
@ 2σ (24-hour time resolution)
3
3
±2 μg/m
< 80 μg/m
; ±5 μg/m3 > 80 μg/m
One minute (updated every 4-seconds)
±0.5 μg/m3 (2-σ, 24-hour time resolution)
0.002% per day (< 0.7% per year)
3
0.1 μg/m
(internally logged and displayed data)
±5% (compared to 24 hour FRM)
Optical: IRLED, 6 mW, 880 nm
Beta: Carbon-14, < 3.7 MBq (< 100 μCi), 5700-year half-life
Optical: silicon/hybrid amplifier
Radiometric: proportional counter (α and β)
3
1 m
/h (16.67 lpm) measured across an internal sub-sonic
orifice
Two serial interface RS232 (25-pin and 9-pin)
4-20mA or 0-10 output of concentration (μg/m
upon order)
*
-22 to 140 °F (-30 to 60 °C)
: Non condensing
100-240V, 50/60Hz, 330W max., 15W without pump or
heater
100-110/100-120V, 50/60Hz or 220/240V, 50/60Hz, 100W
19-inches (W) x 12.25-inches (H) x 13-inches (D) //
483mm(W) x 311mm(H) x 330mm(D)
8.25-inches (W) x 8.75-inches (H) x 4.25-inches (D) /
210mm(W) x 222mm(H) x 108mm(D)
Model 5030 SHARP Monitor Instruction Manual
Introduction
Specifications
3
(auto-ranging)
3
3
) (specify
1-9