Firmware Update Procedure - Thermo Scientific 5030 Instruction Manual

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Firmware Update
Procedure
Thermo Fisher Scientific
Software Requirements
Software "winupdate.exe" for the PC (supplied by Thermo Fisher
Scientific)
Firmware file vx-xxx.hex or MASTER.hex for the instrument
Use the following procedure to update the Model 5030 SHARP firmware.
1. Connect COM1 of the PC to COM1 of Model 5030 and switch on
both instruments.
2. Unlock the keypad in the Operations menu and configure the Model
5030 to:
Print cycle: 0 min 0 sec
Baud rate: 9600 baud
3. Lock the keypad, returning to the remote control mode.
4. Start "winupdate" on the PC.
Note You can check and validate the SHARP firmware version before and
after upload by clicking on the Start button in the Version Query section.
Nephelometer and Beta Attenuation Servicing
Model 5030 SHARP Monitor Instruction Manual
Servicing
7-5