Thermo Scientific 5030 Instruction Manual page 125

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Read Commands
Thermo Fisher Scientific
C
Output of the dust mass concentration in μg/m³ (actual mode)
Output of the β-concentration
C1
C2
Output of the optic concentration
H
Output of the last half hour average in μg/m³
HH
Output of the 1h filtered concentration
H1
1 h average concentration [μg/m³]
H3
3 h average concentration [μg/m³]
HT
24 h (daily) average concentration [μg/m³]
JB
T1, temperature sampling head
JC
T2, temperature dust sampling chamber
JD
T3, temperature inside of the flow meter orifice
JE
T4, temperature external heated sampling tube (at tube surface) all
temperature in °C
JF
P1, differential pressure of the air flow measuring module in Pa
JG
P2, low pressure suction chamber
JH
P3, Barometer in hPa
JI
air flow sampling head (operating flow)
JJ
norm air flow (273K, 1013hPa)
JO
air flow rate regulation 0...100%
JQ
power of external heating 0...100%
JR
relative humidity value (%RH)
m1
non filtered mass in μg
m2
mass on LCD with RC-filter (μg)
m3
mass, filtered with Quick-Lock filter
Remote Control Commands
Model 5030 SHARP Monitor Instruction Manual
RS-232 Commands
B-3

Advertisement

Table of Contents
loading

Table of Contents