Table 6-4. Global Error Status

Code
Global Error Status Definition
01
Sum status of the data and program memory
02
Sum status of sampling and measuring system
04
Sum status of pressure sensors
08
Sum status of air flow regulation
10
Sum status of the sample heater
20
Sum status of temperature measurement
40
Sum status of the Nephelometer measurement
80
Not used
The following break down of the Global Error Status Definitions to a
Detailed Error Status can be obtained by pressing the DISP key once and
the NEXT key four times from the Main Menu.