Thermo Scientific 5030 Instruction Manual page 133

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Remote Control
Thermo Fisher Scientific
Example: GEI (Bavaria/Hessia) protocol including the transmission of the
number of black carbon and the concentration of black carbon:
Inquiry:
<STX>DA<CR>
Response: <STX>MD03
002 +5681+00 00 00 023 000000<SP>
003 +1001+03 00 00 023 000000<CR>
The measuring instrument is controlled by the following protocol:
<STX>ST[Address]<SP>[Command]<ETX>[BCC1][BCC2]
Address: Device address, 3 characters, leading zeros may be replaced by
<SP>.
The address may be also dropped. In this case, the first command character
directly must follow the characters ST.
Command: All commands listed in this appendix are admissible commands
to be applied. The response coming from the device, however, is sent
without protocol. For this reason, no read commands should be used.
In case that the control command is accepted by the measuring instrument
(the characters ST follow the <STX>, correct address, correct block check),
then the command will be executed.
It is allowed to replace <ETX>[BCC1][BCC2] by <CR>.
Data Protocols with Measuring Network Connection
001
+2578
+03 00 00
Address
Conc
Model 5030 SHARP Monitor Instruction Manual
RS-232 Commands
023
000000<SP>
Serial Number
B-11