Installation
Menu Tutorial
2-10
Model 5030 SHARP Monitor Instruction Manual
By pressing NEXT, from the last screen within the main User Screen, the
following screen will appear:
This screen provides the current aerosol
concentration and accumulated mass in
micrograms. Press NEXT.
This screen shows the measured volumetric flow rate
in liters per hour (l/h), the standard flow rate (Nl/h),
as well as the pump controller output in percent of
the output range (such as, 68.6%). Divide by 60 to
obtain liters per minute. Press NEXT.
This screen shows the most recent 30–minute,
one-hour, three-hour, and 24-hour average
concentrations carried over from prior operation.
Press NEXT.
This screen shows the background radon content
measured by the instrument. Press NEXT.
This screen shows the individual text codes in
relation to the combined status codes that are
expected at startup. Press TEXT for additional
information.
This screen shows the additional "POWER ON"
status code after starting the instrument. Press
NEXT.
This screen shows the detailed error status code.
If necessary, this should be interpreted through
the "Troubleshooting" chapter. Press NEXT.
This is the last screen within the main User
Screen, and this shows the current Date
(yy-mm-dd) and Time (hh:mm:ss). Refer to the
"Operation" chapter for changing these values.
Press NEXT.
This is the Operation menu. Press NEXT.