Status Output by
Potential Free Relay
Contacts
Output by Serial
Interfaces (COM1 and
Thermo Fisher Scientific
If the respective status code is not set, the relay circuit will be closed.
If the respective status code is set, the relay circuit will be open.
Each relay has one potential free change contact.
The self-surveillance routine of the Model 5030 SHARP is executed every
second and the relays are set or reset according to the actual status.
Status Reports of the standard relays on the main board include:
Relay 0: signals the "Operation Status"
Relay 1: Nephelometer Auto Ranging
Relay 2: signals the "Error (Break-Down) Status"
Remote Control Commands
After receiving the read command (for example, # <CR>) the device
COM2)
returns:
Data Protocol with Network Connection
Command: <STX> DA address <ETX> <BCC1> <BCC2>
Device returns: <STX> MD 01 address meas.value status SFKT <ETX>
<BCC1> <BCC2>
Refer to "Appendix B" for a detailed description of the code words alone.

Status Output by Potential Free Relay Contacts

Model 5030 SHARP Monitor Instruction Manual
Troubleshooting
6-11