Status Concepts for
Quality-Assured
Measurement
Thermo Fisher Scientific
Status Changes and Entries into the Logbook
When the device status changes, the new device status is copied into the
logbook together with date, time, concentration, mass, air flow rate,
temperatures (T1, T2, T4), RH pressures (P2, P3), and Count Rates (R
R
,).
α
The logbook contains the 20 latest entries (standard). With the option "512
Kbyte RAM-expansion" it has 1632 entries.
Example: (print format no. 11 or no. 19)
Status Definition of the Half-Hour Mean Values of the Concentration
The code[Number]s of all status, occurring during the half-hour, are added
and thus form the status of the half-hour mean values.
This procedure prevents the loss of status information, even when a special
status vanishes again during the half-hour. Multiple occurrence of the same
status in a half hour is not recorded.
Example:
TIME
STATUS
13.45
00 00 00
14.05
00 08 00
14:13
00 00 00
14:35
00 00 40
14:49
00 00 00
14:55
00 00 10
15:10
00 00 00
COMMENTARY
normal operation
relative humidity elevated
normal operation
Nephelometer calibration
stabilizing
normal operation
keypad enabled
normal operation
Model 5030 SHARP Monitor Instruction Manual
Troubleshooting
Print Formats
,
β
6-13