Thermo Scientific 5030 Instruction Manual page 7

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

US NRC Exemption
USEPA PM
EQPM-0609-184
Thermo Fisher Scientific
The Model 5030 SHARP Monitor is an exempt product in accordance
with NRC license No. 20-23922-01E and SSD Registration Certificate
No. AVR-1234-D-101-E. The user is exempt from any licensing
requirements for this device. Laws may vary outside of the United States.
FEM:
Thermo Scientific Model 5030 SHARP Monitor operated for 24-hour
2.5
average measurements; including a 1-micron inlet; PM
connector; sample tube; DHS heater with 35% RH threshold; mass foil kit;
GF10 filter tape; nephelometer zeroing kit; 8-hour filter change; and
operational calibration and servicing as outlined in Model 5030 SHARP
instruction manual.
The following conditions must be followed in order to comply with
USEPA Designation EQPM- 0609-184 for ambient monitoring of PM
1. Operated for 23-25 hour average measurements.
2. Flow rate set to 1000 L/h (16.67 L/min).
3. Automatic filter change (8-hour cycle time).
4. Automatic filter change when flow rate < 950 L/h.
5. Automatic filter change when sampled filter spot mass is > 1,500 μg.
6. DHS set to 35% RH.
7. Calibrated with zero and span mass transfer foil kit for 0 to 5,000
μg/m
8. System calibrated for temperature, barometric pressure, and volumetric
flow rate.
3
range.
VSCC ; inlet
2.5
Model 5030 SHARP Monitor Instruction Manual
Preface Section
US NRC Exemption
:
2.5
vii