Operation
Operation and Service Menus

Main User Screen

3-2
Model 5030 SHARP Monitor Instruction Manual
In the Main User Screen the status code, error code, and aerosol
concentration are found. The concentration value in micrograms per cubic
3
meter (μg/m
) will be updated every 4 seconds. By pressing DISP and
NEXT other values such as aerosol mass, air flow rate, average concentration
values, error status, and time can be displayed. The Main sub-menus are as
follows:
This is the main User Screen showing a
combined STATUS Code, combined ERROR
Code, and PM Concentration, press DISP.
This screen provides the current aerosol
concentration and accumulated mass in
micrograms, press NEXT.
This display shows the volumetric flow at the
inlet, the flow rate at STP conditions (standard
temperature and pressure, 1013 hPa and 25
and the pump power (%). Please note the
standard temperature is adjustable in the
Calibration menu. Press NEXT.
This screen shows the most recent 30–minute,
one-hour, three-hour, and 24-hour average
concentrations carried over from prior operation.
Please note that the 24-hour average is the "daily"
average. Press NEXT.
This screen shows the background radon content
measured by the instrument and is used for
refining the mass measurement. Press NEXT.
This screen shows the individual text codes in
relation to the combined status codes that are
expected at startup, press TEXT for additional
information.