11. Place the rubber ferrule and coupling over the heated sample tube and
thread the coupling into the lower coupling adapter already secured.
APPLY ONLY FINGER TIGHT.
12. At instrument level, properly align and tighten the stainless steel
knurled nut onto the top of the Model 5030.
13. Return to the roof level and firmly tighten the roof flange-coupling
assembly. Caulk around the coupling-to-sample tube interface to avoid
leaks.
14. Attach the ambient temperature radiation shield assembly to the outer
gray plastic sample tube and insert the ambient temperature sensor into
the radiation shield.
15. Attach the inlet assembly onto the sample tube (for example, inlet
adapter, sharp-cut cyclone, PM
16. Attach vertical bracing to the exterior sample tube in at least two lateral
directions that are 90 degrees apart or employ the use of a tripod
attachment as available.
17. Connect the power cord to the rear of the Model 5030 and to a properly
grounded power supply of the appropriate voltage and frequency.
Should the Model 5030 turn ON at this point, set the power switch to
the OFF position and then proceed.
18. Connect the braided vacuum hose to the top brass port on the pump
and the other end to vacuum pump hose connection on the rear of the
monitor. Properly tighten both ends of this braided hose.
19. Connect the 9-pin vacuum pump control cable to the 0-10V Pump
control connector on the rear of the monitor.
20. Connect the vacuum pump power cord to a properly grounded power
supply of the appropriate voltage and frequency.
Inlet).
10
Model 5030 SHARP Monitor Instruction Manual
Installation
Setup and Installation
2-21