Thermo Scientific 5030 Instruction Manual page 116

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Servicing
Nephelometer and Beta Attenuation Servicing
7-8
Model 5030 SHARP Monitor Instruction Manual
The first screen that appears on the SHARP instrument indicates the
address of the data bytes being received. This screen is shown below. The
transfer takes approximately three to five minutes.
RECEIVING FLASH DATA
------------------------------------
2C80
Data Byte Address Screen
After a successful firmware upload the Flash - EEPROM is programmed.
The following screen is displayed.
PROGRAMMING FLASHMEM
------------------------------------
FFC0
Programming Flash Memory Screen
After successfully uploading the new firmware to the Model 5030 SHARP,
the instrument will automatically restart with the new firmware version. You
can now close the WinUpdate program.
The initial screen shown at startup appears for approximately 30 seconds.
This screen contains the firmware version. The version (for example, v 1.02)
should be checked to make sure that it corresponds to the new version.
(This can also be done with "winupdate" by clicking on the Version Query
START button.)
If the firmware upload was faulty or the programming of the Flash -
EEPROM did not work correctly, the instrument starts with the old
firmware version.
Note The parameters of the Model 5030 SHARP are not changed when the
firmware is updated. As a precaution, however, the parameters should be
compared with the values that existed before the update was made. This can
be done by using Print Format 8 before and after the update.