Thermo Fisher Scientific
Code
Symbol
Quantity
1
CONC
Aerosol Concentration
2
MASS
Compensated mass
3
Q-OP
Air flow rate (Volumetric Conditions)
4
Q-N
Air flow rate (Standard Conditions)
5
T1
Ambient temperature
6
T2
Temperature at filter tape
7
T4
Temperature of external tube heater surface
8
RH
Sample Relative Humidity
9
P1
Differential pressure / orifice
10
P2
Vacuum under filter
11
P3
Barometric pressure
12
C-1H
Aerosol Concentration filtered 1 hour
13
M-1H
Compensated mass filtered 1 hour
Note Analog output selections on the display will always show units of
milliamp (mA). Corresponding voltage (V) outputs are achieved by jumper
settings. See the "Installation" chapter.
Press Yes.
Here the operator may define the analog output
range.
User-selected zero point output of 0 mA/0V or 4
mA/2V.
Here the user has chosen a live zero to be a 4 mA
(or 2V) analog output. Press NEXT.
Here the user can choose the analog output of the
default channel by using the +/- keypad. Press
NEXT.
User-selected concentration for the zero point of
the analog output range. Possible settings: -100
3
to 1000 μg/m
. Press NEXT.
Model 5030 SHARP Monitor Instruction Manual
Operation
Operation and Service Menus
Unit
μg/m³
μg
l/h
Nl/h
°C
°C
°C
%
hPa
hPa
hPa
μg/m³
μg
3-11