Operation
Operation and Service Menus
Serial Communication
Print Formats
3-4
Model 5030 SHARP Monitor Instruction Manual
Parameters
Print formats are provided by an output spooler. While printing, the normal
measurement operation will not be interrupted and the measuring
instrument can be operated as usual. Remote Control commands are
possible using the second serial interface during printing out the lists via the
first serial interface. From the internal memory, sufficient information from
the print formats is available. The transmission of all half-hourly mean
concentrations over an entire year, including information on date and status,
requires approximately 632 Kbytes. At a bit transmission rate of 9600, the
transmission process will need approximately 10 minutes. The transmission
of hourly values will take less space and time.
Press NEXT.
Pressing the COM key toggles between the
COM1 and COM2 interface. Choose COM1 or
COM2 to change the serial parameters. Press
NEXT.
The print format indicates which data report will
be issued at the serial interface. You can adjust the
print format using the +/- keys. Press NEXT.
Press PRINT to begin the serial output.
After the set time has elapsed, the selected data
report will output to the serial interface. If set to
0 minutes, the next menu allows a cycle ranging
from 0 to 120 seconds. Printouts will be limited
and repeated to this print cycle. Press NEXT.
This display window allows setting the baud
transmission rate. Choose between 300, 600,
1200, 2400, 4800 and 9600 Baud, using the +/-
keys.
This display window allows a device number to be
applied to the instrument and is important when
multiple monitors are tied into a network on the
same communications line.