Troubleshooting
Status Messages

Detailed Status Messages

6-6
Model 5030 SHARP Monitor Instruction Manual
Examples:
Status
Significance
011
0010
0001
0082
0080
0002
00A9
0001
0008
0020
0080
Note Below are the combinations of potential added status codes and their
breakdown. In the STATUS example above, recognize that the "A" in 00A9
shows a combination of 20 and 80 for the Operating Status. Do not confuse
this with a 02 and 08 for Operating Status. The positioning of the
alphabetical character is key to defining which codes the instrument is
reporting. Being able to recognize these subtleties will allow the user to
easily define and troubleshoot the Model 5030 as necessary. Below is a
matrix of alphanumeric combinations.
Status Codes - Alphanumeric Combinations
0
1
4
5 = 4 + 1
8
9 = 8 + 1
C = 8 + 4
D = 8 + 4 + 1
From the Main Menu, a Global Error Status appears as a 2 hexadecimal
figure.
Table 6-4
indicates the codes and general definitions. Each code
should be interpreted by reviewing the detailed error status codes.
Global Error Status
Manual operation
Mechanical Filter change
Power Supply Switched on
Zeroing (mass)
Filter change (mechanical)
Pump off
Calibration enabled
Poewr Supply Switched on
2
3 = 2 + 1
6 = 4 + 2
7 = 4 + 2 +1
A = 8 + 2
B = 8 + 2 + 1
E = 8 + 4 + 2
F = 8 + 4 + 2 +1