Analog Output Test
Thermo Fisher Scientific
Sensor
Units
°C
T1
°C
T2
°C
T3
°C
T4
P1
hPa
P2
hPa
P3
hPa
RH
%RH
chart recorders connected to the unit and for testing cable connections.
Corresponding voltage output setting will also apply to this test, whereas
4 mA = 2V and 20 mA = 10V. When quitting this menu by pressing either
the BACK or NEXT key, the measured value will be again output to the
analog interface.
Description
Ambient temperature sensor
Sample temperature sensor
Orifice temperature sensor
Heater outer-wall temperature sensor
Orifice differential pressure
Differential vacuum under filter tape
Barometric pressure
Sample relative humidity
In this screen, the user can test the analog signal
output being sent to their external data logger.
By pressing the respective key, 0/4 mA or 20
mA, an equivalent analog output is provided to
the analog channel. This allows for adjusting
Model 5030 SHARP Monitor Instruction Manual
Operation
Service Menu
3-15